SCBA Test Benches
The industry’s leading NFPA-compliant SCBA test bench now has USB communication capability.
Features & Benefits
Performs full readiness testing
Face piece leak check
Test bench meet stringent NFPA standards
Quick Specs
Size (excluding head): Height 14.1 inches; Width 13.5 inches; Depth 16.0 inches
Weight (excluding head) 34 pounds
Operating Temperature 41ºF to 113ºF
Storage Temperature 14ºF to 113ºF
Product Overview
The PosiChek family of SCBA test equipment has been the industry standard for thousands of fire departments, refineries, chemical plants and nuclear power stations worldwide since it was first introduced in 1988. The new Posi 3 USB incorporates a number of hardware and software improvements suggested by our customers to keep pace with increasingly rigorous testing criteria.
NFPA 1852 mandates that all compliant SCBA be tested on a calibrated breathing machine before being put into service, and at least once per year thereafter. Posi 3 USB meets the same stringent NFPA standards as PosiChek3.
Additional Features and Benefits:
Posi 3 USB Performs Full Readiness Testing
Facepiece leak check
Exhalation check valve opening pressure
Static facepiece pressure
Bypass flow rate
Pressure gauge accuracy
Low pressure alarm activation
First stage regulator performance
Breathing resistance at standard (40 LPM)
Breathing resistance at maximum (100 LPM)
Minimum PC System Requirements
Windows XP to Windows 7, 32 or 64 bit (Please call to find out which versions are currently available for 64-bit.)
Pentium 4 processor 2.0 GHz or higher
512 MB Ram
50 Mb of available space on hard drive
USB 2.0 port and CD-ROM Drive
Product Overview
The PosiChek family of SCBA test equipment has been the industry standard for thousands of fire departments, refineries, chemical plants and nuclear power stations worldwide since it was first introduced in 1988. The new Posi 3 USB incorporates a number of hardware and software improvements suggested by our customers to keep pace with increasingly rigorous testing criteria.
NFPA 1852 mandates that all compliant SCBA be tested on a calibrated breathing machine before being put into service, and at least once per year thereafter. Posi 3 USB meets the same stringent NFPA standards as PosiChek3.
Additional Features and Benefits:
Posi 3 USB Performs Full Readiness Testing
Facepiece leak check
Exhalation check valve opening pressure
Static facepiece pressure
Bypass flow rate
Pressure gauge accuracy
Low pressure alarm activation
First stage regulator performance
Breathing resistance at standard (40 LPM)
Breathing resistance at maximum (100 LPM)
Minimum PC System Requirements
Windows XP to Windows 7, 32 or 64 bit (Please call to find out which versions are currently available for 64-bit.)
Pentium 4 processor 2.0 GHz or higher
512 MB Ram
50 Mb of available space on hard drive
USB 2.0 port and CD-ROM Drive
Posi3 USB SCBA Test Bench Order Number
Posi3 USB with low and high pressure (2216 and 4500 psig) manifolds, 250 psig medium pressure
transducer, Enhanced Precision (EP) transducer and standard Posi3 USB SCBA test software
54-56-1117
Posi3 USB with low and high pressure (2216 and 4500 psig) manifolds, 500 psig medium pressure
transducer, Enhanced Precision (EP) transducer and standard Posi3 USB SCBA test software
54-56-1717
Posi3 USB with low and high pressure (2216 and 4500 psig) manifolds, 250 psig medium pressure
transducer, Enhanced Precision (EP) transducer, sound detection system (SDS) and standard Posi3 USB
SCBA test software
54-56-2117
Posi3 USB with low and high pressure (2216 and 4500 psig) manifolds, 500 psig medium pressure
transducer, Enhanced Precision (EP) transducer, sound detection system (SDS) and standard Posi3 USB
SCBA test software
54-56-2717
Posi3 USB with low and high pressure (2216 and 4500/5500 psig) manifolds, 250 psig medium pressure
transducer, Enhanced Precision (EP) transducer, to test 5500 psi SCBA models and standard Posi3 USB
SCBA test software
54-56-5117
Posi3 USB with low and high pressure (2216 and 4500/5500 psig) manifolds, 250 psig medium pressure
transducer, Enhanced Precision (EP) transducer, to test 5500 psi SCBA models, sound detection system
(SDS), and standard Posi3 USB SCBA test software
54-56-6117
Test Bench Upgrades Order Number
Add USB capability with advanced software package to a standard PosiChek3 54-21-09*
Add USB capability with advanced software package to a standard PosiChek3 with (SDS) sound detection 54-21-10*
Add 5500 psig upgrade
Note: Unit must be Posi3 USB for 5500 upgrade.
54-21-12*
Add Sound Detection System (SDS) capability
Note: Unit must be Posi3 USB for upgrade. Calibration fee and return shipping not included.
54-21-A17
*Note: includes calibration fee and return shipping anywhere within the US. Contact HA for further details.
Accessories Order Number
Pressure reducing regulator and valve assembly 35-0470
2216 psi Adapter for pressure reducing regulator and valve assembly 35-0475
Replacement test head for Posi3 USB (included in complete PosiChek unit) 54-370
Low pressure exterior manifold, 2216/3000 psi (included in complete PosiChek unit) 54-21-A03
High pressure exterior manifold, 4500/5500 psi (included in complete PosiChek unit) 54-21-A02
Hose assembly, first stage (included in complete PosiChek unit) 35-751
Case for PosiChek, large, foam lined. Holds PosiChek unit, head, and external manifolds
Note: Not recommended for calibration only shipping.
54-21-A15
Case for PosiChek, small, foam lined. Holds only the PosiChek unit
Note: Recommended for calibration only shipping.
54-21-A16
Calibration Services Order Number
Annual Service and Calibration includes return shipping within the USA 54-21-A10-CT22*
Service Expedite fee POSI_RUSH_SVC
*Note: The PosiChek Annual Service Fee includes return shipping anywhere in the United States. PosiCheks shipped outside of the United States are subject
to additional freight, duty and tax fees and are subject to export laws. The PosiChek annual Service Fee does not cover the cost of instrument upgrades,
nor does it cover damage or defects caused by abuse, improper packaging, alterations, or unauthorized repairs. Outside the U.S. market, additional freight
charges will apply.
Posi3 USB Software Order Number
Honeywell (Survivair) OEM software 54-58-01
MSA OEM software 54-58-07
Draeger- US OEM software 54-58-02
Interspiro- US OEM software 54-58-05
Avon/ISI OEM software 54-58-06
Honeywell (Fenzy) OEM software 54-58-04
Standard/Generic software (included with PosiChek unit) 54-58-10
Replacement USB software
Note: an SCBA test record must be provided.
Contact HA
Note: Scott Safety software must be purchased directly from Scott Safety.
PosiChek3 32-bit software Order Number
Honeywell (Survivair) OEM software 32-bit 54-22-69-L
MSA OEM software 32-bit 54-22-65-L
Draeger OEM software 32-bit 54-22-58-L
Interspiro-US OEM software 32-bit 54-22-62-L
Avon/ISI OEM software 32-bit 54-22-64-L
Standard/Generic software 32-bit 54-22-66-L
Replacement 32-bit software
Note: an SCBA test record must be provided.
Contact HA
Note: Scott Safety software must be purchased directly from Scott Safety
Operating Ranges
1. Power Requirements
Line Voltage range 100 VAC to 125 VAC (use 2A input fuses); or
220 VAC to 250 VAC (use 1A input fuses).
Line Voltage selection Fuse change required; otherwise it is automatic.
Line Frequency 50 to 60 Hz
Maximum Power Consumption 60VA
2. Environmental Characteristics
Ambient Temperature Ranges
Operating 41º F to 113º F (5º C to 45º C)
Storage 14º F to 113º F (-10º C to 45º C)
Pollution Degree 2
3. Physical Characteristics
Size (excluding head)
Height 14.1 inches (358 mm)
Width 13.5 inches (343 mm)
Depth 16.0 inches (406 mm)
Weight (excluding head) 34 pounds (15.4 kg)
4. Fuse requirements
120 VAC Systems 2A, Slow Blow, 250 VAC, 5 x 20 mm
220 VAC Systems 1A, Slow Blow, 250 VAC, 5 x 20 mm